3 Installation
IC equipment: Dialysis
■■■■■■■■
7
■ Loosen the screws with the hex key.
■ Remove the screws with the washers and put them aside.
3
Disassembling the dialysis cell
■ Remove the upper part of the dialysis cell.
■ Remove the sealing ring.
4
Cleaning the dialysis cell
CAUTION
Damage to the dialysis cell
Organic solvents (e.g.: acetone) corrode and damage the dialysis
cell material (PMMA).
■ Use only ultrapure water or a water-ethanol mixture (70:30) for
cleaning the dialysis cell.
■ For samples that contain organic components (e.g. solvents),
use the PEEK dialysis cell (6.2729.120). This cell has an excellent
chemical resistance to organic chemicals.
■ Thoroughly rinse off the sealing ring (2-3), the left chamber (2-2)
and the right chamber (2-4) of the dialysis cell with ultrapure
water.
■ Dry all parts with a lint-free cloth.