Metrohm IC Net 2.0 Uživatelský manuál Strana 84

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 196
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 83
6 Devices
IC Net 2.0
76
Second column Program instruction (see below).
Third column Parameter for program instruction (see
below).
ENABLED Enable program start (a disabled program
is not started).
<Add> Add new program instruction.
<Delete> Delete selected program instruction.
<Verify> Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time pro-
gram on the Program page:
Instruction Parameter entry Meaning
Program RESET, END Set program flag. The END flag
can be used to end a program,
especially if the program time
should be longer than the chro-
matogram duration. Additional
steps after this flag are not al-
lowed. The RESET flag is used to
reset the parameters to the sys-
tem startup values.
Pump on, off Switch on or off the pump drive.
Links
The Links tab of the 752 Pump Unit is used for COM port selection
and settings (details see section 5.2.4).
Zobrazit stránku 83
1 2 ... 79 80 81 82 83 84 85 86 87 88 89 ... 195 196

Komentáře k této Příručce

Žádné komentáře