6.8 Metrohm solvent delivery unit
IC Net 2.1
83
Program
The
Program
tab of the
SDU
window contains a time program for
the pump control system to define eluent gradients.
Flow
Overall flow of pump control system in
mL/min. This value is also available for the
WATCH WINDOW
. This field is only available
if the
Show flow column
option is disabled.
Entry range:
0.01 ... 20.00 mL/min
(depending
on number of pumps)
Stop pumps after the run
Stop pumps after the determination.
Program table
Number
Row number in the program table (read-
only).
Time
Time at which the flow and/or gradient is
changed.
Entry range:
0.0 ... 99999 min
Flow
Overall flow of pump control system in
mL/min.
Entry range:
0.01 ... 20.00 mL/min
(depending
on number of pumps)
%B
Individual flow rate for pump B in percent of
total flow.
Entry range:
0 ... 100 %
%C
Individual flow rate for pump C in percent of
total flow.
Entry range:
0 ... 100 %
%D
Individual flow rate for pump D in percent of
total flow.
Entry range:
0 ... 100 %
Zobrazit stránku 90