6 Devices
IC Net 2.3
214
End time
Subtab for the end time settings of the
timebase program.
Current Displays the current End time of the Triath-
lon autosampler.
Method value Manual setting of the end time. The end
time is sent to the Triathlon Autosampler
with
<SET>.
Entry range: 0:00:00…9:59:59 h:mm:ss
auto time
enabled Auto setting of the end time is enabled, the
end time is equal to the analysis time (see
Methods programming/Triathlon meth-
ods/Injection/Injection
) if the end of a time-
base program exceeds the analysis time, it
is overruled.
disabled The end time specified under Method value
is selected, if a timebase program exceeds
the end time, the programmed events after
the end time are not carried out during the
run.