6.5 761 Compact IC
IC Net 2.3
89
761 unit version Selection of instrument version:
761.0010 = 761 Compact IC without sup-
pressor
761.0020 = 761 Compact IC with suppres-
sor
Suppressor Suppressor module:
Autostep with Automatic switching to the next position
each time the injection valve is switched ei-
ther to the Fill or to the Inject position.
Use peristaltic pump
If this option is disabled, the peristaltic
pump is not switched on automatically at
the start of a determination or with Startup
hardware
.
Polarity Selection of the output signal polarity:
+ positive polarity (for anions)
negative polarity (for cations)
Measure channel Display of the data source selected in the
Data source window (see section 6.26.2).
Links
The Links tab of the 761 Compact IC is used for COM port selection
and settings (details see section 5.2.4 Links).