2 Remote control commands
780/781 pH/Ion Meter - Operation via RS232 11
│ │ │ │ Bec4, Bec7, Bec10, Rad1.09, Rad1.68,
│ │ │ │ Rad4.01, Rad6.84, Rad7.00, Rad7.38,
│ │ │ │ Rad9.18, Rad10.01, Bak4, Bak7, Bak9,
│ │ │ │ Bak10, Ham4.01, Ham7.00, Ham9.21,
│ │ │ │ Ham10.01, Pre4, Pre7, Pre9, Own1, Own2,
│ │ │ │ Own3, Own4, Own5
│ │ │ ├ .2 Buffer 1
│ │ │ │ └ .Select Selection of buffer 1 Met7, same select. as for buffer 1
│ │ │ ├ .3 Buffer 1
│ │ │ │ └ .Select Selection of buffer 1 Met9, same select. as for buffer 1
│ │ │ ├ .4 Buffer 1
│ │ │ │ └ .Select Selection of buffer 1 NIST1, same select. as for buffer 1
│ │ │ └ .5 Buffer 1
│ │ │ └ .Select Selection of buffer 1 NIST4, same select. as for buffer 1
│ │ ├ .Limits Limits for calibration data
│ │ │ ├ .SlopeMin Lower limit for slope [%] 0.1...95.00...999.9
│ │ │ ├ .SlopeMax Upper limit for slope [%] 0.1...103.0...999.9
│ │ │ ├ .OffsetMin Lower limit for pH (0) 0.0...6.400...99.999
│ │ │ └ .OffsetMax Upper limit for pH (0) 0.0...8.000...99.999
│ │ └ .UOffset Acquisition of measured values $G
│ │ ├ .Status Status of UOffset ON, OFF
│ │
└ .Value UOffset [mV] -2200.0...0.0...2200.0
│ ├ .LimitspH
│ │ ├ .Status Status limits pH ON, OFF
│ │ ├ .UpperLim Upper limit pH -19.999...14.000...19.999
│ │ ├ .UHystereses Upper hysteresis pH -19.999...0.020...19.999
│ │ ├ .LowerLim Lower limit pH -19.999...0.000...19.999
│ │ └ .LHystereses Lower hysteresis pH -19.999...0.020...19.999
│ ├ .LimitsT
│ │ ├ .Status Status limits T ON, OFF
│ │ ├ .UpperLim Upper limit T [°C] -999.9...100.0...999.9
│ │ ├ .UHystereses Upper hysteresis T [°C] -999.9...0.2...999.9
│ │ ├ .LowerLim Lower limit T [°C] -999.9...0.0...999.9
│ │ └ .LHystereses Lower hysteresis T [°C] -999.9...0.2...999.9
│ ├ .PlotPara Parameters for live plot on a printer
│ │ ├ .LeftMargin Left margin of pH scale -19.9...0.0...19.9
│ │ ├ .RightMargin Right margin of pH scale -19.9...14.0...19.9
│ │ ├ .TLeftMarg Left margin of temperature scale [°C] -999...20...999
│ │ └ .TRightMarg Right margin of temperature scale [°C] -999...30...999
│ ├ .Presel Preselections
│ │└ .IReq Request of identifications after start id1, id1 & id2, OFF
│ ├ .ElTestPara Electrode test parameters
│ │ ├ .Type Type of pH electrode standard, gel, non-aq., own
│ │ ├ .ElExc Excellent electrode
│ │ │ ├ .ElParam Electrode parameters
│ │ │ │ ├ .StreamPot Streaming potential [mV] -999.9...2.5...999.9
│ │ │ │ ├ .Drift Drift [mV/min] 0.1...2.0...9.9
│ │ │ │ ├ .MinSlope Minimum slope [%] 0.1...96.50...999.9
│ │ │ │ ├ .MaxSlope Maximum slope [%] 0.1...101.0...999.9
│ │ │ │ ├ .Response Response time [s] 0...45...9999
│ │ ├ .ElGood Good electrode
│ │ │ ├ .ElParam Electrode parameters
│ │ │ │ ├ .StreamPot Streaming potential [mV] -999.9...3.0...999.9
│ │ │ │ ├ .Drift Drift [mV/min] 0.1...2.5...9.9
│ │ │ │ ├ .MinSlope Minimum slope [%] 0.1...96.00...999.9
│ │ │ │ ├ .MaxSlope Maximum slope [%] 0.1...102.0...999.9
│ │ │ │ ├ .Response Response time [s] 0...50...9999
│ │ ├ .ElPass Excellent electrode
Komentáře k této Příručce