2 Remote control commands
780/781 pH/Ion Meter - Operation via RS232 19
&Setup
Object Description Input range
&Setup Settings for the operating mode
: ├ .Keycode Send key code ON, OFF
├ .Trace Message on changed values ON, OFF
├ .Lock Lock key functions
│ ├ .Keyboard Lock all keyboard keys ON, OFF
│ ├ .Config Lock <CONFIG> key ON, OFF
│ ├ .Parameter Lock <PARAM> key ON, OFF
│ ├ .Cal Lock <CAL> key ON, OFF
│ ├ .UserMeth Lock user method functions
│ │ ├ .Recall Lock recall function ON, OFF
│ │ ├ .Store Lock store function ON, OFF
│ │ └ .Delete Lock delete function ON, OFF
│ ├ .CalData Lock <CAL DATA> key ON, OFF
│ ├ .Mode Lock <MODE> key ON, OFF
│ └ .ElTest Lock <EL.TEST> key ON, OFF
├ .AutoInfo Automatic message for status changes
│ ├ .Message Definition of message
│ │ └ .DateTime Date and time of occurrence ON, OFF
│ ├ .P When mains is switched on ON, OFF
│ ├ .G When method id started ON, OFF
│ ├ .R When "ready" ON, OFF
│ ├ .S When stopped ON, OFF
│ ├ .E Error ON, OFF
│ ├ .Re Request after start ON, OFF
│ ├ .D When drift condition OK ON, OFF
│ ├ .I Change upon inputs ON, OFF
│ └ .O Change upon outputs ON, OFF
├ .InputAssign Assignment of I/O line inputs
│ ├ .pH Mode pH 0...1...15
│ ├ .T Mode T 0...2...15
│ ├ .U Mode U 0...3...15
│ ├ .pHcal Start pH calibration 0...5...15
│ ├ .ElTest Start electrode test 0...6...15
│ ├ .Conc Mode Conc 0...8...15
│ ├ .ConcCal Start concentration calibration 0...9...15
│ └ .Enter <ENTER> key 0...15
└ .Graphics Changing graphics output
├ .Grid Grid on curve ON, OFF
├ .Frame Frame around curve ON, OFF
├ .Recorder Length of axis
│ ├ .Right Length of measured value axis 0.4...0.8...1.0
│ └ .Feed Length of paper drive axis 0.4...1.0
└ .InstrNo Device identification $G
└ .Value Input of device identification 8 ASCII char.